Singapore recorded a record 3.22 billion gross tonnage of vessel arrivals in 2025. For a ship on that route, port time is limited, and an instrument that fails calibration at the wrong moment costs time. This article explains what needs calibration on a ship, what changed in 2025 for gas detectors, and how to plan calibration around a call at Singapore.
What Needs Calibration
Which instruments apply depends on the ship type, so treat the list as a starting point. Where no convention sets an interval, the maker’s instructions do.
|
Instrument |
Why it matters |
Basis |
|
Portable and personal gas detectors |
Testing the atmosphere before entering an enclosed space. Detectors should be calibrated and tested, and measure CO2 as well as oxygen, flammable and toxic gases |
SOLAS XI-1/7; Res. MSC.581(110) |
|
15 ppm bilge alarm (oil content meter) |
Calibrated by the maker or an authorised person at intervals not exceeding five years, or the maker’s term if shorter. The certificate stays on board |
MEPC.107(49) as amended; IACS UI MPC127 |
|
Fixed gas and oxygen analysers |
Span, zero and functional checks on vapour emission control and inert gas systems |
Maker instructions |
|
Pressure gauges, transmitters and pressure calibrators |
Calibration against traceable references |
Maker instructions |
|
Thermometers and temperature calibrators |
Temperature calibration and verification |
Maker instructions |
|
Sounding (UTI) tapes |
Inspection, calibration and repair |
Maker instructions |
|
Load cells |
Calibration and verification of load-measuring equipment |
Maker instructions |
The CO2 Change for Enclosed Space Entry
IMO adopted Resolution MSC.581(110) on 27 June 2025. It took effect on 3 December 2025 and replaced Resolution A.1050(27). CO2 must read below 0.5% by volume, or 5,000 ppm, before entry. Detectors should measure CO2 as well as oxygen, flammable gases and toxic gases, and people entering a space should carry calibrated and tested personal detectors. A standard four-gas instrument may not meet this unless it also measures CO2.
The Resolution is a recommendation, while SOLAS regulation XI-1/7 still requires at least two sets of portable gas detection equipment. It also recommends managing entry within the safety management system and keeping a ship-specific register of enclosed spaces.
Calibration Is Not a Bump Test
A bump test checks that an instrument responds to gas and that its alarms work. Calibration adjusts the instrument against a certified reference gas and is recorded on a certificate. Crews bump test as the maker advises. Calibration is done on a schedule by a competent provider. Both matter, and neither replaces the other .
Bilge Alarms and ODME
The calibration certificate for a 15 ppm bilge alarm is checked at annual, intermediate and renewal surveys under the International Oil Pollution Prevention Certificate. The five-yearly calibration does not have to coincide with the renewal survey, so it can be planned at the most convenient port. Oil discharge monitoring equipment (ODME) is inspected and calibrated to the maker’s instructions.
Instruments Ships Bring Most Often
These are the instruments we calibrate most often, in order of how frequently they come to us:
- Pressure calibrators
- Temperature calibrators
- UTI sounding tapes
- Gas detectors
- 15 ppm bilge alarm monitoring devices
ODME, pressure gauges and load cells follow close behind. The work is done onboard or at our service station, whichever suits the call.
Planning Calibration Around a Singapore Call
- List every instrument with its next due date before the ship arrives.
- Check which gas detectors measure CO2, and plan upgrades early.
- Order the matching calibration gas and confirm the certificate you need.
- Hand over instruments early, or arrange work onboard, and ask for as-found and as-left readings.
If instruments fall due between ports, plan the work at the call with the longest stay. Marine Mechanics tracks due dates and notifies you before they expire, not after a finding.
What the Certificate Should Show
It should show the as-found and as-left readings, the traceable reference used, and the next due date. That documentation protects the vessel’s record at inspection.
How Marine Mechanics Supports Ships Calling Singapore
Marine Mechanics operates from five locations, including Singapore. Its calibration scope covers:
- Personal, portable and fixed gas detectors: inspection, calibration, bump testing and certification.
- Upgrades from four-gas to CO2-capable five-gas instruments.
- Fixed vapour and inert gas oxygen analysers.
- 15 ppm bilge alarms and ODME.
- Pressure calibrators and gauges, temperature calibrators and thermometers, UTI sounding tapes and load cells.
- pH meters and nitrogen generator instruments.
We also supply detectors, sensors and certified calibration gas blends with regulators, including matching gas for CO2-capable five-gas detectors.
Three Quick Questions
Do gas detectors on ships have to measure CO2?
IMO’s recommendations, in effect since 3 December 2025, say detectors should measure CO2, and set the entry limit below 0.5% by volume.
How often must a 15 ppm bilge alarm be calibrated?
At intervals not exceeding five years, or the maker’s term if shorter, with the certificate kept on board.
How many portable gas detection sets must a ship carry?
At least two, under SOLAS regulation XI-1/7.
Marine Mechanics is a technical engineering partner for the full vessel lifecycle. To plan calibration around your next Singapore call, write to mmsafety@marinemechanics.biz or see our Calibration Services page.